HT instruments GEO416 User Manual
Page 23

GEO416
-
GEO416GS
EN - 22
9.1.1. Reference
standards
Safety: IEC/EN61010-1,
IEC/EN61557-1, IEC/EN61557-5
Measuring accessories’ safety:
IEC/EN61010-031
Technical literature: IEC/EN61187 
Insulation: double 
insulation
Mechanical protection:
IP50 according IEC/EN60529
Pollution level:
2
Measurement category:
CAT III 240V (to earth), maximum 415V between inputs
Max height of use:
2000m (6562ft)
 
9.1.2. General 
features
Mechanical features 
Dimensions (L x W x H): 
235 x 165 x 75mm (9 x 6 x 3in)
Weight (including batteries):
1kg (2lv)
 
Power supply 
Battery type: 
6x1.5V batteries type AA LR6 or
6x1.2V rechargeable batteries NI-MH AA LR6 2100mA
Low battery indication:
symbol “
” is displayed
Battery life:
approx. 500 tests
Auto Power OFF:
after 3 min of idleness
 
Display 
Features: 
LCD custom with back-light 73x65 mm
 
Memory 
Features: 
999 memory locations
 
PC connection 
Features: 
optoinsulated port for bi-directional communication
 
9.2. ENVIRONMENT 
9.2.1. Operating 
environmental
conditions
Reference calibration temperature: 23 ± 5°C (73 ± 41°F) 
Working temperature: 
0 ÷ 40°C (32°F ÷ 104°F)
Maximum relative humidity:
<80%RH
Storage temperature:
-10 ÷ 60°C (14°F ÷ 140°F)
Storage humidity:
<80%RH
This instrument complies with the requirements of the European Low Voltage
Directive 2006/95/CE (LVD) and EMC Directive 2004/108/CE
9.3. ACCESSORIES 
9.3.1. Standard and optional accessories GEO416 
See enclosed packing list 
 
9.3.2. Standard accessories GEO416GS 
 Set of 4 cables, banana-banana, L=1m
KIT416CV
 Set of 4 alligator clips
COC4-UK
Carrying
bag BORSA2000N
 ISO9000 calibration certificate
 User
manual