 - GROUND RESISTIVITY MEASUREMENT
The ground resistivity value is an essential parameter to calculate the resistance value of 
the earth rods to be used for the earth installation’s construction. The measurement is 
effected according to standards IEC 781, VDE 0413, IEC/EN61557-5. 
 
CAUTION
 The instrument can be used for voltage and current measurements on
installations with over voltage category equal to CAT III 240V to earth and 
maximum voltage of 415V between inputs. Do not connect the instrument 
to installations whose voltages exceed the limits indicated in this manual. 
Exceeding such limits may cause electric shock to the user and damage 
the instrument 
 Always connect the cables to the instrument and to the alligator clips when
the latter are not connected to the plant under test
 Always respect the Hand-held area of probe (see § 4.2)
 If the length of the supplied cables isn’t suitable for the plant under test (see
§ 11), You can create your own extensions following indications in § 11.2.1
Fig. 6: Ground resistivity measurement
1.
Turn on the instrument pressing the ON/OFF key
2.
Pressing right/left arrow keys ,  select MOD, then pressing up/down arrow 
keys ,  select 
 option