The s rod’s resistance r, Ω] (where r, Is the earth resistance value) – HT instruments GEO416 User Manual
Page 16: The h rod’s resistance r

GEO416
-
GEO416GS
EN - 15
4.5.1. Anomalous measuring applications
1.
When starting a measurement the 
instrument checks the continuity of 
measuring cables. If the volt circuit 
(red cable S and green cable ES) is 
interrupted or its resistance value is 
too high, the instrument displays a 
screen similar to the one beside. Check 
that terminals are properly connected 
and that the earth rod is connected to 
Volt circuit’s resistance too 
high 
Value of input interfering 
voltage 
Example for 3W mode 
 
terminal S and not driven into a pebbly or scarcely conductive ground. In this latter 
case pour water around the rod to decrease its resistance value (§ 11.2). 
RP>top is displayed when: 
The S rod’s resistance R
S
> 50KΩ is summed up to the volt circuit
The resistance of rod S exceeds the value 1200 + 100 R
X
[Ω] (where R
X
is the
earth resistance value)
2.
When starting a measurement the 
instrument checks the continuity of 
measuring cables. If the ampere 
circuit (blue cable H and black cable 
E) is interrupted or its resistance 
value is too high, the instrument 
displays a screen similar to the one 
beside. Check that terminals are 
properly connected and that the earth 
rod is connected to 
Ampere circuit’s resistance 
too high 
Value of input interfering 
voltage 
 
Example for 3W mode 
terminal H and not driven into a pebbly or scarcely conductive ground. In this latter 
case pour water around the rod to decrease its resistance value (§ 11.2). 
RC>top is displayed when: 
The H rod’s resistance R
H
> 50KΩ is summed up to the ampere circuit
The resistance of rod H exceeds the value 1200 + 100 R
X
[Ω] (where R
X
is the
earth resistance value)
3.
When starting a measurement the 
instrument checks the continuity of 
measuring cables. If the volt circuit 
(red cable S and green cable ES) and 
the ampere circuit (blue cable H and 
black cable E) are both interrupted or 
their resistance values are too high, 
the instrument displays a screen similar 
to the one beside. Check that the 
Both volt and ampere 
circuits’ resistance too high
Value of input interfering 
voltage 
 
Example for 3W mode
terminals are properly connected and that the earth rods connected to terminals S 
and H are not driven into a pebbly or scarcely conductive ground. In this latter case 
pour water around the rods to decrease their resistance value (§ 11.2). 
RP, RC>top is displayed when: 
The S rod’s resistance R
S
> 50KΩ is summed up to the volt circuit and the H
rod’s resistance H R
H
> 50KΩ is summed up to the ampere circuit
Both the S rod’s resistance and the H rod’s resistance exceed the value 1200 +
100 R
X
[Ω] (where R
X
is the earth resistance value)
 
 