En - 16, The above said anomalous cases cannot be stored – HT instruments GEO416 User Manual
Page 17

GEO416
-
GEO416GS
EN - 16
4.
When starting a measurement, if the 
red cable (connected to S terminal) 
and the green cable (connected to ES 
terminal) are reversed, the instrument 
do not perform the test, emits a long 
sound tone and displays the screen 
beside 
Red and green cables 
reversed 
 
Example under ρ mode
5.
When starting a measurement, if an 
interfering voltage higher than 9V is 
detected at the volt circuit’s input, 
the instrument do not perform the test, 
emits a long sound tone and displays 
the screen beside 
Too high input interfering 
voltage at volt circuit
Input interfering voltage 
value 
Example under 2W mode
6.
When starting a measurement, if an 
interfering voltage higher than 9V is 
detected at the ampere circuit’s 
input, the instrument do not perform the 
test, emits a long sound tone and 
displays the screen beside 
Too high input interfering 
voltage at ampere circuit 
Input interfering voltage 
value 
 
Example under 2W mode
7.
When starting a measurement, if an 
interfering voltage higher than 9V is 
detected at both the ampere and volt 
circuits’ inputs, the instrument do not 
perform the test, emits a long sound 
tone and displays the screen beside 
Too high input interfering 
voltage at both ampere 
and volt circuits
Input interfering voltage 
value 
Example under 2W mode
8.
If battery voltage is too low the 
instrument displays the symbol of low 
battery as well as the message 
and no measurement is allowed. It is 
however possible to carry out settings, 
reading of stored data, etc 
Too low power supply, low 
batteries 
Input interfering voltage 
value and distance set 
among rods 
Example under ρ mode
9.
The above said anomalous cases cannot be stored 
 