Ohaus NBF1A2 Navigator Balances User Manual
Page 94

31
TM
RS232 COMMAND TABLE (Cont.)
Command
Character
Description
x%
Set % reference function. Uses x (Real Number) as current Percent
Reference. * Reference weight must be entered in grams.
%
Print current Percent Reference Weight. *
x#
Set PC reference function. Uses x (Real Number) as current Part
Reference. * Reference weight must be entered in grams.
#
Print current Parts Count Reference Weight. *
F
Print current function.
FV
Print final configuration version
V
Print software version
*
Balance must be in correct weigh mode.
?
Print current mode
mg
g
kg
dwt
ct
oz
oz t
GN
tael (Hong Kong)
tael (Singapore)
tael (Taiwan)
momme
lb
N
tical
Pcs
%
Field: Mode CR
LF
Length:
5
1
1
NG
Same as pressing G/N/T button
Z
Same as pressing zero button.
T
Same effect as pressing Tare button.
xD
Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON)
xI
Set Averaging Filter Level to “x”,
where x = 0 to 3 (see table).
0
=
minimum level
1
=
2
=
3
=
maximum level
TIME
Print Current Time. Note, a ? mark will follow if date or time has not been set.
hh:mm:ss SETTIME Set Time Command and Remove Invalid Time Indicator
DATE
Prints Current Date. Note, a ? mark will follow the year if date or time has
not been set.
mm/dd/yy SETDATE Set Date Command and remove Invalid Indicator
xT
Download tare value, tare weight must be entered in grams.
SN
Show Serial Number.
SW
Show Lockswitch status.
LE
Show Last Error Code. Response: Err: Error Number.
R$TM
Read Clock Trim value.
W$TM
Write Clock Trim Value. The clock can be adjusted by +60 seconds a day.
Global Setting List command
Resets Setup and Print menus to factory defaults. CAUTION: This will reset
RS232 configuration.