Command
Character
Description
xZ
Set Auto Zero to “x”,where x = 0 to 3 ). 0=Off, 1=0.5d, 2=1d, 3=5d.
x%
Set % reference function. Uses x (Real Number) as current Percent Reference.
*
Reference weight must be entered in grams.
x#
Set PC reference function. Uses x (Real Number) as current Part Reference.
*
Reference weight must be entered in grams.
Esc R
Resets Setup and Print menus to factory defaults.
CAUTION: This will reset RS232 configuration.
ON
Turns balance on.
OFF
Turns balance off.
#
Print current Parts Count Reference Weight.
%
Print current Percent Reference Weight.
xAW
Set Animal Level from 0 to 3. 0= least amount of sampling.
xE
Set/Reset Auto Restart in Animal mode. Where x is 0=Off and 1=ON.
E
Start Animal cycle.
xT
Download tare value, tare weight must be entered in grams.
ID
Print Current User ID String.
xID
Program User ID String, 1-8 characters.
AC
Abort Calibration.
xUC
Starts User Weight Calibration. Input format: x is the user weigh value and must be
at least 50% of full span value. (enter full numerical value with no decimal point).
LE
Show Last Error Code. Response: Err: Error Number.
SN
Show Serial Number.
xS
Print Stable Only. Where x =0 Off and x=1 On.
TIME
Print Current Time. Note, a ? mark will follow if date or time has not been set.
mm/dd/yy SETDATE Set Date Command and remove Invalid Indicator
hh:mm:ss SETTIME Set Time Command and Remove Invalid Time Indicator
DATE
Prints Current Date. Note, a ? mark will follow the year if date or time has not been
set.
W$TM
Write Clock Trim Value. The clock can be adjusted by +60 seconds a day.
R$TM
Read Clock Trim value.
SW
Show Lockswitch status.
xCN
Set Checkweighing nominal weight. Reference weight must be entered in grams.
*
xCO
Set Checkweighing over limit. Reference weight must be entered in grams.
*
xCU
Set Checkweighing under limit. Reference weight must be entered in grams.
*
CN
Print Checkweighing nominal weight.
CO
Print Checkweighing over limit.
CU
Print Checkweighing under limit.
xAO
Set Auto-shutoff. - Set timer from 0-60 minutes. 0 = timer disabled.
AO
Print Auto Shutoff
RS232 COMMAND TABLE (Cont.)
*