3.3.3 Calibration Adjust

Balances with InCal

TM

contain software which allows the

internal calibration value to be adjusted by + 100 divisions
at full scale capacity using an external mass which is
traceable to a certified standard.

Procedure

• Perform the internal calibration procedure.

• Press >O/T< button to zero the balance.

• Place a mass equal to the

span calibration value of

the balance on the pan. Note the reading on the bal-

ance and see if the balance indicates the exact

weight or indicates a higher or lower reading. If the

reading is higher or lower, proceed.

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYP is displayed.

• Press

button until , CAL ADJ is displayed.

• Press Enter button, CAL ADJ 0 should be dis-

played (0 is factory setting).

NOTE: Balance will retain last CAL ADJ setting.

• Press

or

button until desired number is

displayed.

• Press Enter button, SAVED is momentarily displayed,

then display indicates LOCK.

• Press Enter button, LOCK ON or LOCK OFF is

displayed.

• Press

or

button and select either LOCK ON

or LOCK OFF.

• Press Enter button, SAVED is momentarily displayed,

then EXIT is displayed.

• Press Enter button, display returns to WEIGH

mode. Note weight reading and remove mass from

pan.

• Perform the internal calibration procedure. The value

entered as an adjustment is now stored. Place the

calibration mass on the pan and check. Repeat pro-

cedure if further correction is required.

• To return to factory setting, follow procedure above

and set CAL ADJ to 0.