Visara UCT-L User Manual
Page 18

1-10
707108-001
UCT-L User’s Manual
Electrical interference and atmospheric considerations
Electrostatic discharge in the vicinity of the unit should be minimized by avoiding high
resistance floor material and carpeting that does not have anti-static properties, by
avoiding the use of plastic seats and covers, and by avoiding low humidity levels. The
unit should be located away from areas that generate electromagnetic interference (for
example, transformers, power distribution panels, welding equipment, motors, and
transmitters). Do not put the unit next to fluorescent lights. The UCT-L should not be
installed where the atmosphere contains corrosive elements that can damage the unit.
Do not run the power cable or signal cables in areas that produce electromagnetic
interference (for example, near transformers, switching equipment, welding equipment,
power distribution panels, and under carpets where vacuum cleaning is done). Also, do
not roll heavy equipment over the cable.