Analog output, Ieee 488, Caution – Measurement Computing DAC488 v.1 User Manual
Page 9: Tput…… 3 ieee 488…… 3
![background image](https://www.manualsdir.com/files/797763/content/doc009.png)
DAC488 User’s Manual
DAC488 Overview 3
Analog Output
WARNING
WARNING
WARNING
WARNING
There may be up to 200 V isolation potential between analog common and earth (CE
version) or up to 500 V isolation potential (Standard version). Avoid contact with
analog output terminals and the cables connected to them. Do not apply power until
after the AC line and analog output connections have been properly made. The AC
line must remain connected to the unit whenever the analog output terminals are
connected to a device exceeding 60 VDC or 30 Vrms common mode voltage. Contact
with voltage potentials could cause bodily injury or death!
CAUTION
CAUTION
CAUTION
CAUTION
Solid wire is recommended for the analog output terminal connections. If stranded
wire is used, then strip the wire 6 mm, and twist or tin the stripped end. After
tightening the connector, check that no loose wire strands are present.
Terminal Installation Category: Standard: Not Applicable. CE: Category 1.
DC Output Voltage and Resolution
1V Range:
±
1.02375 V, 250 µV/bit
5V Range:
±
5.11875 V, 1.25 mV/bit
10V Range:
±
10.2375 V, 2.5 mV/bit.
DC Output Current and Output Resistance: 10 mA maximum; and <500 m
Ω
(typical)
Accuracy (25
±±±±
5
°°°°
C; Iout = 1 mA):
1V Range:
±
(0.05% of setting + 1 mV)
5V Range:
±
(0.05% of setting + 3 mV)
10V Range:
±
(0.05% of setting + 10 mV)
Warm Up Time to Rated Accuracy: 1 hour
Temperature Coefficient (0
°°°°
to 20
°°°°
C, and 30
°°°°
to 50
°°°°
C):
±
(0.002% of setting + 100 µV) /
°
C
Channel to Channel Isolation: Standard: 500 V max, 100,000 V-Hz.
CE: 200 volts max, 100,000 V-Hz
Channel to Digital Low Isolation: Standard: 500 V max, 100,000 V-Hz.
CE: (Channel Common to Earth) 200 V max, 100,000 V-Hz
Connectors: One 12-pin quick disconnect analog.
IEEE 488
CAUTION
CAUTION
CAUTION
CAUTION
The IEEE 488 terminal must only be used to control a non-isolated IEEE 488 system.
The common mode voltage (cable shell to earth) must be zero.
Terminal Installation Category: Standard: Not Applicable. CE: Category 1.
Implementation: SH1, AH1, T4, TE0, L4, LE0, SR1, RL0, PP0, DC1, DT1,E1.
Terminators: Programmable CR, LF, LF CR and CR LF with EOI.
Programmable: IEEE Bus Terminators, EOI Control, SRQ Mask, Port Voltage, Gain, Offset, Digital I/O
lines, Output Format, Internal Buffer, Mode.
Connector: Standard IEEE 488 connector with metric studs.