Automatic calibration summary report – Measurement Computing DAC488 v.1 User Manual
Page 56

50 DAC488 Calibration
DAC488 User’s Manual
After the last port is tested, a summary report will be printed similar to that below:
Calibration Report
Manufacturer: Model: DAC488/4 Serial#
Port
Range
%Err +FS
%Err -FS
Offset Err
J
Vals
H Val
1
1
+0.003
-0.003
+28.000E-06
136
146
+15
Passed
1
2
-0.002
+0.000
-20.000E-06
118
126
+17
Passed
1
3
+0.001
-0.001
+0.000E+00
116
123
+18
Passed
2
1
-0.004
+0.003
-37.000E-06
112
98
-1
Passed
2
2
-0.001
+0.000
+0.000E+00
120
104
+3
Passed
2
3
-0.002
-0.001
-20.000E-06
119
102
+5
Passed
3
1
-0.001
+0.001
-27.000E-06
115
111
+11
Passed
3
2
-0.003
+0.001
-20.000E-06
118
113
+8
Passed
3
3
-0.001
+0.001
+0.000E+00
112
106
+1
Passed
4
1
+0.003
-0.007
+41.000E-06
123
113
+8
Passed
4
2
-0.002
+0.003
-50.000E-06
122
111
+18
Passed
4
3
+0.001
-0.001
-10.000E-06
127
114
+30
Passed
Automatic Calibration Summary Report
A description of each item in the auto-calibration summary report is given below:
•
Port: This column shows which of the four ports is being tested.
•
Range: This column shows which of the three volt ranges is being used, where 1 represents the
±
1 volt
range, 2 represents the
±
5 volt range, and 3 represents the
±
10 volt range.
•
%Err +FS: This is the percent of Error at positive Full Scale (FS) Reading. This is computed using
the following formula:
(FS Reading - FS Voltage)/(FS Voltage) x 100% = %Err +FS
For example, if a DAC port is programmed for 10 volts on the 10 volt scale and the voltmeter reading
10.0001 volts, then the percent of Error would be:
(10.0001 - 10.0000)/(10.0000) x 100% = +0.001%
•
%Err - FS: This is the percent of Error at negative Full Scale (FS) Reading. This is computed using
the following formula:
(FS Reading - FS Voltage)/(FS Voltage) x 100% = %Err -FS
For example, if a DAC port is programmed for -5 volts on the 5 volt scale and the voltmeter reading -
5.0002 volts, then the percent of Error would be:
(-5.0002 - -5.0000 )/( -5.0000 ) x 100% = -0.004%
•
Offset Err: This column shows the measured output voltage when the port is programmed to output 0
volts.
•
J Vals: These two columns show the positive and negative Gain Calibration Constants for each port
and range. The positive Gain Calibration Constant is printed first.
•
H Val: This column shows the Offset Calibration Constant for each range and port.