Part iii: automatic calibration report, Automatic calibration report format – Measurement Computing DAC488 v.1 User Manual
Page 55

DAC488 User’s Manual
DAC488 Calibration 49
Part III: Automatic Calibration Report
After the automatic calibration finishes, a calibration report will be printed if the DAC488 is connected to
an IEEE printer. The following example shows the calibration report of the three volt ranges for Port 1:
DAC488/4
Performance Verification
Port # 1
Nominal
Actual
Error
Result
Nominal
Actual
Error
Result
10.0000
10.0000
0.0000
Pass
-10.0000
-10.0000
0.0000
Pass
9.0000
8.9979
0.0021
Pass
-9.0000
-8.9976
0.0024
Pass
8.0000
7.9997
0.0003
Pass
-8.0000
-7.9994
0.0006
Pass
7.0000
7.0003
0.0003
Pass
-7.0000
-7.0001
0.0001
Pass
6.0000
6.0011
0.0010
Pass
-6.0000
-6.0009
0.0009
Pass
5.0000
4.9999
0.0001
Pass
-5.0000
-4.9997
0.0003
Pass
4.0000
4.0004
0.0004
Pass
-4.0000
-4.0002
0.0002
Pass
3.0000
2.9998
0.0003
Pass
-3.0000
-2.9996
0.0004
Pass
2.0000
2.0005
0.0005
Pass
-2.0000
-2.0004
0.0004
Pass
1.0000
0.9993
0.0007
Pass
-1.0000
-0.9992
0.0008
Pass
0.0000
0.0000
0.0000
Pass
-0.0000
-0.0000
0.0000
Pass
5.0000
5.0000
0.0000
Pass
-5.0000
-5.0000
0.0000
Pass
4.5000
4.4989
0.0011
Pass
-4.5000
-4.4988
0.0012
Pass
4.0000
3.9998
0.0003
Pass
-4.0000
-3.9997
0.0003
Pass
3.5000
3.5002
0.0002
Pass
-3.5000
-3.5001
0.0001
Pass
3.0000
3.0005
0.0005
Pass
-3.0000
-3.0005
0.0005
Pass
2.5000
2.4999
0.0001
Pass
-2.5000
-2.4999
0.0001
Pass
2.0000
2.0001
0.0001
Pass
-2.0000
-2.0002
0.0002
Pass
1.5000
1.4999
0.0001
Pass
-1.5000
-1.4998
0.0002
Pass
1.0000
1.0002
0.0002
Pass
-1.0000
-1.0002
0.0002
Pass
0.5000
0.4996
0.0004
Pass
-0.5000
-0.4996
0.0004
Pass
0.0000
0.0000
0.0000
Pass
-0.0000
-0.0000
0.0000
Pass
1.0000
1.0001
0.0001
Pass
-1.0000
-1.0000
0.0000
Pass
0.9000
0.8998
0.0002
Pass
-0.9000
-0.8997
0.0003
Pass
0.8000
0.8000
0.0000
Pass
-0.8000
-0.7999
0.0001
Pass
0.7000
0.7001
0.0001
Pass
-0.7000
-0.7000
0.0000
Pass
0.6000
0.6001
0.0001
Pass
-0.6000
-0.6000
0.0000
Pass
0.5000
0.5000
0.0000
Pass
-0.5000
-0.4999
0.0001
Pass
0.4000
0.4001
0.0001
Pass
-0.4000
-0.3999
0.0001
Pass
0.3000
0.3000
0.0000
Pass
-0.3000
-0.2998
0.0002
Pass
0.2000
0.1995
0.0005
Pass
-0.2000
-0.1993
0.0007
Pass
0.1000
0.0997
0.0003
Pass
-0.1000
-0.0995
0.0005
Pass
0.0000
0.0000
0.0000
Pass
-0.0000
-0.0000
0.0000
Pass
Automatic Calibration Report Format
A report similar to the one above will be printed for all four ports. A description of each item in the auto-
calibration report is given below:
•
Nominal: The voltage the DAC was programmed to output.
•
Actual: The actual voltage measured by the DMM.
•
Error: The difference between the Nominal and the Actual values.
•
Result: Indicates Pass if the port is functioning within the stated specifications.