beautypg.com

Part iii: automatic calibration report, Automatic calibration report format – Measurement Computing DAC488 v.1 User Manual

Page 55

background image

DAC488 User’s Manual

DAC488 Calibration 49

Part III: Automatic Calibration Report

After the automatic calibration finishes, a calibration report will be printed if the DAC488 is connected to
an IEEE printer. The following example shows the calibration report of the three volt ranges for Port 1:

DAC488/4

Performance Verification

Port # 1

Nominal

Actual

Error

Result

Nominal

Actual

Error

Result

10.0000

10.0000

0.0000

Pass

-10.0000

-10.0000

0.0000

Pass

9.0000

8.9979

0.0021

Pass

-9.0000

-8.9976

0.0024

Pass

8.0000

7.9997

0.0003

Pass

-8.0000

-7.9994

0.0006

Pass

7.0000

7.0003

0.0003

Pass

-7.0000

-7.0001

0.0001

Pass

6.0000

6.0011

0.0010

Pass

-6.0000

-6.0009

0.0009

Pass

5.0000

4.9999

0.0001

Pass

-5.0000

-4.9997

0.0003

Pass

4.0000

4.0004

0.0004

Pass

-4.0000

-4.0002

0.0002

Pass

3.0000

2.9998

0.0003

Pass

-3.0000

-2.9996

0.0004

Pass

2.0000

2.0005

0.0005

Pass

-2.0000

-2.0004

0.0004

Pass

1.0000

0.9993

0.0007

Pass

-1.0000

-0.9992

0.0008

Pass

0.0000

0.0000

0.0000

Pass

-0.0000

-0.0000

0.0000

Pass

5.0000

5.0000

0.0000

Pass

-5.0000

-5.0000

0.0000

Pass

4.5000

4.4989

0.0011

Pass

-4.5000

-4.4988

0.0012

Pass

4.0000

3.9998

0.0003

Pass

-4.0000

-3.9997

0.0003

Pass

3.5000

3.5002

0.0002

Pass

-3.5000

-3.5001

0.0001

Pass

3.0000

3.0005

0.0005

Pass

-3.0000

-3.0005

0.0005

Pass

2.5000

2.4999

0.0001

Pass

-2.5000

-2.4999

0.0001

Pass

2.0000

2.0001

0.0001

Pass

-2.0000

-2.0002

0.0002

Pass

1.5000

1.4999

0.0001

Pass

-1.5000

-1.4998

0.0002

Pass

1.0000

1.0002

0.0002

Pass

-1.0000

-1.0002

0.0002

Pass

0.5000

0.4996

0.0004

Pass

-0.5000

-0.4996

0.0004

Pass

0.0000

0.0000

0.0000

Pass

-0.0000

-0.0000

0.0000

Pass

1.0000

1.0001

0.0001

Pass

-1.0000

-1.0000

0.0000

Pass

0.9000

0.8998

0.0002

Pass

-0.9000

-0.8997

0.0003

Pass

0.8000

0.8000

0.0000

Pass

-0.8000

-0.7999

0.0001

Pass

0.7000

0.7001

0.0001

Pass

-0.7000

-0.7000

0.0000

Pass

0.6000

0.6001

0.0001

Pass

-0.6000

-0.6000

0.0000

Pass

0.5000

0.5000

0.0000

Pass

-0.5000

-0.4999

0.0001

Pass

0.4000

0.4001

0.0001

Pass

-0.4000

-0.3999

0.0001

Pass

0.3000

0.3000

0.0000

Pass

-0.3000

-0.2998

0.0002

Pass

0.2000

0.1995

0.0005

Pass

-0.2000

-0.1993

0.0007

Pass

0.1000

0.0997

0.0003

Pass

-0.1000

-0.0995

0.0005

Pass

0.0000

0.0000

0.0000

Pass

-0.0000

-0.0000

0.0000

Pass

Automatic Calibration Report Format

A report similar to the one above will be printed for all four ports. A description of each item in the auto-
calibration report is given below:

Nominal: The voltage the DAC was programmed to output.

Actual: The actual voltage measured by the DMM.

Error: The difference between the Nominal and the Actual values.

Result: Indicates Pass if the port is functioning within the stated specifications.

This manual is related to the following products: