To verify the port voltage – 10-volt range, To verify the port voltage – 5-volt range, Range…… 21 – Measurement Computing DAC488 v.1 User Manual
Page 27
DAC488 User’s Manual
DAC488 Power & Assembly 21
To Verify the Port Voltage – 10-Volt Range
1. Send the following command string to the DAC488:
C0 P1 A0 R3 V10 X
where:
•
C0
= Direct Control Mode
•
R3
= 10-Volt Range
•
P1
= Port 1
•
V10
= Output +10 Volts
•
A0
= Autorange Off
•
X
= Execute
2. Send the following command string to the 199 DMM:
F0 R3 Z0 N1 S1 X
where:
•
F0
= DC Volts
•
N1
= Scanner Channel 1
•
R3
= 30-Volt Range
•
S1
= 5 ½ Digit Resolution
•
Z0
= Zero Disabled
•
X
= Execute
3. For each test, the specific Value Output (
V
) command, its commanded output voltage, and its
corresponding Low Limit and High Limit, are as follows:
•
V10
= Output +10 V
+9.985
≤
Reading
≤
+10.015 Volts
•
V5
= Output +5 V
+4.990
≤
Reading
≤
+5.010 Volts
•
V1
= Output +1 V
+0.994
≤
Reading
≤
+1.006 Volts
•
V0
= Output 0 V
-0.005
≤
Reading
≤
+0.005 Volts
•
V-1
= Output -1 V
-1.006
≤
Reading
≤
-0.994 Volts
•
V-5
= Output -5 V
-5.010
≤
Reading
≤
-4.990 Volts
•
V-10
= Output -10 V
-10.015
≤
Reading
≤
-9.985 Volts
To Verify the Port Voltage – 5-Volt Range
1. Send the following command string to the DAC488:
C0 P1 A0 R2 V5 X
where:
•
C0
= Direct Control Mode
•
R2
= 5-Volt Range
•
P1
= Port 1
•
V5
= Output +5 Volts
•
A0
= Autorange Off
•
X
= Execute
2. Send the following command string to the 199 DMM:
F0 R3 Z0 N1 S1 X
where:
•
F0
= DC Volts
•
N1
= Scanner Channel 1
•
R3
= 30-Volt Range
•
S1
= 5 ½ Digit Resolution
•
Z0
= Zero Disabled
•
X
= Execute
3. For each test, the specific Value Output (
V
) command, its commanded output voltage, and its
corresponding Low Limit and High Limit, are as follows:
•
V5
= Output +5 V
+4.9920
≤
Reading
≤
+5.0080 Volts
•
V2.5
= Output +2.5 V
+2.4945
≤
Reading
≤
+2.5055 Volts
•
V1
= Output +1 V
+0.9960
≤
Reading
≤
+1.0040 Volts
•
V0
= Output 0 V
-0.0030
≤
Reading
≤
+0.0030 Volts
•
V-1
= Output -1 V
-0.9960
≤
Reading
≤
-1.0040 Volts
•
V-2.5
= Output –2.5 V
-2.4945
≤
Reading
≤
-2.5055 Volts
•
V-5
= Output -5 V
-4.9920
≤
Reading
≤
-5.0080 Volts