beautypg.com

Specifications – Hoefer IEF100 User Manual

Page 8

background image

• p2

Specifications


Capacity using
Single Electrode set

1– 6 IPG strips

7 to 24 cm long

Capacity using
Dual Electrode accessory

2– 12 IPG strips

7 cm long

Output

Volt

12,000 V

Current

10 mA

Power

10 W

Power in

100 – 240 V
50 – 60 Hz
100 W

User interface

Large blue/white graphical display

Knob

7 button keypad

Platform temperature

15 – 25 °C

Protocol capacity

30 programs with 9 steps each

Input/output ports

Ethernet, RS232

Dimension (w × d × h)

38 cm × 27 cm × 19 cm

Weight

8 kg

Environmental

Indoor use

4 – 40 °C

operating conditions

Humidity

up to 80%

Altitude

up to 2000 m

Installation category

II

Pollution degree

2

Product certifications

EN61010-1:2001,

EN61326:1998, CE, WEEE, RoHS

This declaration of conformity is only
valid for the instrument when it is:

• used in laboratory locations,

• used as delivered from Hoefer, Inc.

except for alterations described in the
user manual, and

• connected to other CE-labeled instru-

ments or products recommended or
approved by Hoefer, Inc.