Run screen – Hoefer IEF100 User Manual
Page 22

• p16
Run Screen
The information on the run screen is described in detail in the IEF
operation section.
The run screen has no editable fields. It does have the following four
soft keys, MAIN, PAUSE, NEXT and GRAPH (Fig. 12).
Note: The IEF100 instrument monitors
the current and wattage in each
individual strip. If necessary, conditions
can be adjusted and/or problematic
strips can be removed.
MAIN
Allows the user to review and edit all protocols while the current IEF
is still running.
PAUSE
The PAUSE button interrupts the high voltage to the IPG strips,
allowing the user to interface safely with the strips.
After a pause, RUN will continue from where it was interrupted, or
ABORT will terminate the run.
NEXT
The NEXT button allows the user to skip the remaining portion of
the active step, and start the next step in the protocol.
GRAPH
The GRAPH button will display the current run results graphically.
Use the knob to scroll through the voltage profile, and the μA profiles
in channels 1– 6.
Press EXIT to return to the run screen.
At the end of the run the IEF100 will beep, and turn off the high
voltage power supply. A screen displays the protocol number, the end
date and time, and the final IEF conditions in each IPG strip.
Fig. 12. Run screen.
Note: The IEF100 also beeps
between step changes.