Hoefer IEF100 User Manual
Page 40

• p34
When focusing is complete, the IEF100 will beep, and an end run
screen will be displayed (Fig. 36).
Line 1
The protocol number and name. The word finished is displayed in the
upper right corner.
Line 2
The number of strips run (or number of channels if using the Dual Electrode
Accessory).
Line 3
The date and time that focusing terminated.
Line 4
The total time and volt-hours of focusing.
Line 5
The last step time and volt-hours.
Line 6
The volts, and the total current and watts of all IPG strips at the end
of the run.
Line 8–13
The conditions in each strip at the end of the run.
The results can be viewed graphically by pushing the function GRAPH.
The data will be retained in memory until a new run is started.
At this point, the IPG strips can be stored at -20 °C.
Or, if the second dimension is to proceed immediately, the buffer
needs to be exchanged in a two step equilibration process before
second dimension PAGE.
Note: Whenever a new IEF is
started, the previous IEF data is
permanently lost.
Note: Do not equilibrate the IPG
strips before storing at -20 °C.
Fig. 36. End run screen.