• p15

Step Time

Steps can be programmed to run for a set amount of time, Hrs (00:00
in hours:minutes), or for a set amount of volt-hours, Vhrs (volts
multiplied by the number of hours). Most IEF strip manufacturers
recommend to run the IPG strips to a specific volt-hours endpoint.

The researcher should determine the endpoint for their specific sample.

The first step with a 0 value entered in the step time is treated as the
end of the protocol.

Step Value

Knob

Computer

Settable

Settable

Step Type

Range

Increments Increments

Constant Volt

0 – 12,000

250

1

Gradient Volt

0 – 12,000

250

1

Constant Watt

0.1 – 2.0

0.1

0.1


Step Time

Hrs (hours:minutes)

00:00 – 99:59

0:15

0:01

Vhrs (volt-hours)

0 – 300,000

500

1

Note: When switching between time
units, the IEF100 will convert between
hours and Vhrs automatically. It may
help to set the step time units (Hrs or
Vhrs) before the numerical endpoint
is entered.