CMG-DM24

Whilst the file is loading, a progress window will be displayed.
Depending on the speed of the link, it may take up to 20 minutes to

transfer the firmware.

Once the file is fully transferred, the DM24 will return to the
maintenance menu. Press ENTER, and wait whilst the system boots.

You can run system code other than the default by pressing R at the

maintenance menu, followed by the slot number of the new code. This
will run the DM24 with the new firmware for one session only. When

it is rebooted, the DM24 will revert to the Default application.

Loading new DSP code

Press D at the maintenance menu. The digitizer will reply with

Enter 0/1 to select DSP code to update

Select which of the two DSP code slots you want to overwrite, and
press ENTER. The default is always 0.

Enter Filename/date – upto 31 characters

You can enter a descriptive string for the DSP code here. The DM24
will print this string at every boot-up, to remind you which version of

the DSP code you are using.

When you press ENTER, the digitizer will then request a transfer using
the Xmodem protocol, as described above.