3 general configuration – Guralp Systems CMG-DM24 User Manual
Page 53

Operator's guide
•
The command may be in the extended dictionary. This
dictionary contains certain advanced commands, as well as
exposing internal digitizer functions and variables. To enable
the extended dictionary, issue the command ok-1 (that's a
number one, not a lower-case 'L').
When you have finished using the extended dictionary, close it
with the command [seal] (including the square brackets), so
you do not inadvertently change the internal settings with
subsequent commands.
•
Alternatively, the command may not be implemented in the
firmware version you have installed. Check the Güralp Systems
web site for updated firmware.
EXPLAIN
Syntax: EXPLAIN command
Displays a short message describing the syntax of command and its
effect.
5.3 General configuration
SET-ID
Syntax: SET-ID (interactive)
Sets the system identifier and serial number of the DM24 to values you
supply.
SET-ID
System Identifier ? {ALPHA} DM24
Serial # ? (C23300) C90200
The system identifier you supply may contain up to 5 alphanumeric
(0-9A-Z) characters.
The serial number you supply must contain exactly 6 alphanumeric
(0-9A-Z) characters and should end with two zeroes (0) in almost all
circumstances. The first four characters are copied verbatim into
stream IDs and the final two characters are used as the “base” for
generating stream identifiers. All stream IDs used in this manual
assume that these two characters are zeroes.
October 2009
53