Guralp Systems CMG-DM24 User Manual
Page 39
![background image](https://www.manualsdir.com/files/788438/content/doc039.png)
Operator's guide
STA and LTA periods, together with the Ratio, without restarting the
digitizer (see Section 4.2, page 44).
Since it is not generally advisable to trigger from broadband data, the
digitizer provides a set of standard bandpass filters to apply to the data
streams before they are tested for the trigger condition. This filtering
serves to maximise sensitivity within the frequency band of interest,
and filter out noise outside this band. You can select which bandpass
filter to use from the Bandpass filter drop-down menu. The corner
frequencies of the pass band of the filter are determined by the Nyquist
frequency, which is derived from the sampling rate of the triggering
data. The three filter options have pass bands between 10% and 90%,
between 20% and 90% and between 50% and 90% of the data’s
Nyquist frequency, respectively.
The possible filter configurations are shown in the following table:
Rate (samples/s) Bandwidth 1 (Hz)
Bandwidth 2
(Hz)
Bandwidth 5
(Hz)
1000
50 – 450
100 – 450
250 – 450
500
25 – 225
50 – 225
125 – 225
400
20 – 180
40 – 180
100 – 180
250
12.5 – 112.5
25 – 112.5
62.5 – 112.5
200
10 – 90
20 – 90
50 – 90
125
6.25 – 56.25
12.5 – 56.25
31.25 – 56.25
100
5 – 45
10 – 45
25 – 45
50
2.5 – 22.5
5 – 22.5
12.5 – 22.5
40
2 – 18
4 – 18
10 – 18
25
1.25 – 11.25
2.5 – 11.25
6.25 – 11.25
20
1 – 9
2 – 9
5 – 9
10
0.5 – 4.5
1 – 4.5
2.5 – 4.5
8
0.4 – 3.6
0.8 – 3.6
2 – 3.6
5
0.25 – 2.25
0.5 – 2.25
1.25 – 2.25
4
0.2 – 1.8
0.4 – 1.8
1 – 1.8
2
0.1 – 0.9
0.2 – 0.9
0.5 – 0.9
1
0.05 – 0.45
0.1 – 0.45
0.25 – 0.45
October 2009
39