AMETEK XG Family Programmable DC User Manual
Page 51

Local
Operation
M370430-01 Rev E
3-5
3
To access the tracking mode for entering voltage and current:
1.
Select the VOLTS or AMPS position on the 9-position mode
control. If the set point is blinking, the unit is in coarse
tracking mode.
•
When the VOLTS mode is selected, the voltage set point will
blink in the output voltage display.
•
When the AMPS mode is selected, the current set point will
blink in the output current display.
2.
Use the rotary Adjust/Enter control to adjust the set point.
Note that there is an activity time out (see Normal Display
Mode and Inactivity Timeout on page 3-6).
The adjustments you make to the set point affect the opposing
output value (voltage when in AMPS mode or current when in
VOLTS mode), which will update in the non-blinking display.
3.
Press the rotary Adjust/Enter control to enter fine adjust
tracking mode.
The set point blinks faster when the unit is in fine adjust tracking
mode.
4.
Use the rotary Adjust/Enter control to fine tune the set point.
5.
Once the set point has been selected, press the rotary
Adjust/Enter control to exit tracking mode and return to
normal display mode.
To access the select and set entry mode for the voltage and current
set points:
1.
Select VOLTS or AMPS position on the 9-position mode
control.
2.
Press the rotary Adjust/Enter control three times in
succession to enter coarse Pre-Set Adjustment.
3.
Press the rotary Adjust/Enter control one more time to enter
fine Pre-Set Adjustment.
4.
Press the rotary Adjust/Enter control once again to set the
value.
The units can also timeout (adjustable) to exit the menu. Once
Pre-Set Mode is entered, CrPU or FnPU will appear in the voltage
display.