Resistive-controlled voltage isol setup – AMETEK XG Family Programmable DC User Manual
Page 128

Analog
Programming (APG) and Isolated Analog Programming (ISOL)
4-28
M370430-01 Rev E
Resistive-Controlled Voltage ISOL Setup
Activating ISOL Programming Voltage Mode
To activate ISOL programming voltage mode using an external
resistor:
1.
Turn the 9-position mode control to the VAP position or press
the rotary adjust/Enter control if the control knob is already at
the VAP position.
is displayed on the output voltage display.
2.
Turn the rotary Adjust/Enter control to select
(resistor
source from isolated connector).
3.
Press the rotary adjust/Enter control to commit the selected
mode.
is displayed on the output voltage display.
4.
Select the input source level using the rotary adjust/Enter
control.
5.
Once the desired level has been selected, press the rotary
Adjust/ Enter control to commit the change.
The display will return to the normal display.
Deactivating ISOL Voltage Mode
To deactivate the ISOL voltage mode:
1.
Turn the 9-position mode control to the VAP position or press
the rotary Adjust/Enter control if the control knob is already
at the VAP position.
2.
is displayed on the output voltage display.
3.
Turn the rotary Adjust/Enter control to select
.
4.
Press the rotary adjust/Enter control to commit the change.
The display will return to the normal display.