Voltage-controlled current isol setup – AMETEK XG Family Programmable DC User Manual
Page 125

Analog Programming
(APG) and Isolated Analog Programming (ISOL)
M370430-01 Rev E
4-25
4
Voltage-Controlled Current ISOL Setup
Activating ISOL Programming Current Mode
1.
Turn the 9-position mode control to the CAP position or press
the rotary adjust/Enter control if the control knob is already at
the CAP position.
is displayed on the output voltage display.
2.
Turn the rotary Adjust/Enter control to select
(voltage
source from isolated connector).
3.
Press the rotary adjust/Enter control to commit the selected
mode.
is displayed on the output voltage display.
4.
Select the input source level using the rotary adjust/Enter
control.
5.
Once the desired level has been selected, press the rotary
adjust/ Enter control to commit the change.
The display will return to the normal display.
Deactivating ISOL Current Mode
To deactivate the ISOL current mode:
1.
Turn the 9-position mode control to the VAP position or press
the rotary Adjust/Enter control if the control knob is already
at the VAP position.
2.
is displayed on the output voltage display.
3.
Turn the rotary Adjust/Enter control to select
.
4.
Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.