Storing calibration data, Step 3: recalibrate gain – AMETEK XG Family Programmable DC User Manual
Page 269

Calibration and
Troubleshooting
M370430-01 Rev E
7-5
7
Step 3: Recalibrate Gain
M
e
a
s
u
re
d
V
a
lu
e
Programmed Value
Programmed Value
Figure 7-4
Calibration: Step 3 Recalibrate Gain
Repeat Step 1 for best results.
Storing Calibration Data
Calibration data will not be stored until a CAL:STOR command is
issued. Without calling this command, the new calibration will be lost
after a power cycle.
If desired, the unit may be returned to its calibration state when it
shipped from the factory. To do this, the CAL:REST command may be
issued.
Important:
sThe CAL:REST command will cause the nonvolatile memory of
the supply to be overwritten with the factory calibration state.
Command
Description
[:]CALibration:STORe
Saves calibration data to nonvolatile memory
[:]CALibration:RESTore Resets the calibration to the Factory state
M
e
a
s
u
re
d
V
a
lu
e