beautypg.com

AMETEK SLD Series User Manual

Page 74

background image

Remote Programming

Sorensen SLD-Series DC Load

5-20

M540075-01 Rev C

FALL Time

Purpose:

The dynamic load mode's FALL slew rate query command.

Command Syntax:
FALL?{NL}

Description:
FALL? returns the presently programmed low load current level in Constant Current mode
The engineering unit is “A/us”.

SHORT ON/OFF

Purpose:
SHORt ON or OFF status query command.

Command Syntax:
SHOR?{NL}

Description:
SHOR? returns the presently SHORt status, “0” indicates SHORt OFF, and “1” indicates SHORt ON.

DYNAMIC ON/OFF

Purpose:
DYNamic ON or OFF status query command

Command Syntax:
DYN?{NL}

Description:
DYN? returns the presently DYNamic ON or OFF status, “0” indicates static load mode or DYNamic
OFF, and “1” indicates DYNamic load mode or DYNamic ON.