Remote Programming
Sorensen SLD-Series DC Load
5-10
M540075-01 Rev C
STOR{SP}{m[,n]}{;|NL}
Description:
Parameter m is 1~5 for 5 different states with electronic load’s status and load current into the non-
volatile memory.
Parameter n is 1-30 for 30 memory bank for 150 (m*n) different state with DC electronic load
module's load status and load current into memory in the electronic loads.
Example:
STORE 2,30; store the DC electronic load module's load status and load current into the memory
147.
RECALL
Purpose:
Recall the state of load level and status, is stored by the GPIB/RS232 STORe command.
Command Syntax:
REC{SP}{m[,n]}{;|NL}
Description:
This command is used to recall the memory state, which is stored into memory by the GPIB/RS232
STORe command. Up to 150 states can be recalled for DC electronic load modules.
Example:
REC 2,30; Recall the state of load level and status that is stored in memory 147 by GPIB/RS232
STOR command.
REC 147; Recall the state of load level and status that is stored in memory 147 by GPIB/RS232
STOR command.
VOLTAGE Limit
Purpose:
To set the upper/lower limit value of threshold voltage.
Command Syntax:
LIM:VOLT:{HIGH|LOW}{SP}{NR2}{;|NL}
Description:
This command is to set the upper/lower limit value of threshold voltage. When input voltage is lower
than the lower limit value or higher than the upper limit value, NG indicating light will come on to
indicate “NO GOOD."
Example:
LIM:VOLT:LOW 1.0 ; To set the lower limit value of threshold voltage to 1.0 V.
LIM:VOLT:HIGH 200.0 ; To set the upper Limit vale of threshold voltage to 200.0V.
CURRENT Limit
Purpose:
To set the upper/lower limit value of threshold current.
Command Syntax:
LIM:CURR{HIGH/LOW}{SP}{NR2}{;|NL}
Description: