Sorensen SLD-Series DC Load

4-2

M540075-01 Rev C

4.2

Remote Sense Connections

Figure 4-2 illustrates a typical set up with the electronic load connected for remote sense
operation. The remote V-sense BNC cable of the electronic load are connected to the output of
the power supply. Remote sensing compensates for the voltage drop in applications that
require long lead lengths.

The 5-character voltage Meter of SLM-series DC electronic load measures the voltage of V-
sense BNC input Terminal automatically, so the high accuracy 5-character voltage Meter can
measure the specific points voltage of the power supply's output voltage.

Load leads should be bundled or tie wrapped together to minimize inductance.

Figure 4-2 Remote Voltage Sense Connections