1 operating modes – AMETEK SLD Series User Manual

Page 15

background image

Sorensen SLD-Series DC Load

Introduction

M540075-01 Rev C

1-3

1.1.1 Operating

Modes

The operating modes of the SLD-series Electronic Load include Constant Current (CC) mode,
Constant Resistance (CR) mode, and Constant Voltage (CV) mode. The wide range dynamic
load with rise/fall current slew rate is available in CC mode only.

CC Mode:

(Constant Current mode)

With the operating mode of constant current, the SLD-series Electronic load will sink a
current in accordance with the programmed value regardless of the input voltage (see
Figure 1-5).

I

V

CC

LOAD
CURRENT

CURRENT SETTING

INPUT VOLTAGE

Figure 1-5 Constant Current Mode

CR Mode:

(Constant Resistance Mode)

At constant resistance mode; the SLD-series Electronic Load will sink a current linearly
proportional to the load input voltage in accordance with the programmed resistance
setting (see Figure 1-6).

I

V

LOAD
CURRENT

INPUT VOLTAGE

RESISTANCE
SETTING

Figure 1-6 Constant Resistance Mode