3 cv mode, 4 dynamic mode – AMETEK SLD Series User Manual
Page 41
Sorensen SLD-Series DC Load
Operation
M540075-01 Rev C
3-13
3.3.3 CV
Mode
CV mode operates in Static mode only.
1. Press the MODE key (3) until the CV LED (3) is lit.
2. Press the PRES key (9) to view the programmed values.
3. To set the voltage level, use the
ÕÖ
keys (13) to select each place-value to
change, then press the
×
Ø
keys (13) to set to the desired voltage level.
4. To exit the preset mode, press the PRES key until its LED is not lit.
5. Press the LOAD key to toggle the load off or on.
• Change the voltage at any time using the
×
Ø
keys.
• View the value set by pressing the PRES key.
3.3.4 Dynamic
Mode
Dynamic mode allows fast state switching in either current or power mode. The dwell time
and current slew rate are controlled in addition to the high and low values.
1. Press the MODE key (3), until the CC LED (3) is lit.
2. Set high and low values
a. Press the DYN STTING key (12) until its LED is lit to view the programmed
values. The high level must be higher than the low level; in setting the values
below, the load will not allow the HIGH level to be lower than the LOW level.
b. To set the high level, press the DYN STTING key until “CurrH” appears on the
upper display. The lower display shows its value.
c. Use
the
ÕÖ
keys (13) to select each place-value to change, and press the × Ø
(13) keys to set to the desired high current level.