AMETEK SLD Series User Manual
Page 69
Sorensen SLD-Series DC Load
Remote Programming
M540075-01 Rev C
5-15
LOAD ON Voltage
Purpose:
LOAD ON voltage level query command.
Command Syntax:
LDON?{;|NL}
Description:
LDON? returns the present load on voltage of SLH-series electronic load module.
LOAD OFF Voltage
Purpose:
LOAD OFF voltage level query command.
Command Syntax:
LDOF?{;|NL}
Description:
LDOF? returns the present load off voltage of SLH-series electronic load module.
LEVEL HIGH/LOW
Purpose:
Static mode's LEVEL low or high status query command or active LEVEL of AC electronic load
query command.
Command Syntax:
LEVE?{NL}
Description:
LEVE? returns the present level status, “0” indicates LEVEl LOW, and “1” indicates LEVEl HIGH.
PRESET ON/OFF
Purpose:
PRESet ON or OFF status query command.
Command Syntax:
PRES?{NL}
Description:
PRES? returns the present preset status, “0” indicates PRESet OFF, and “1” indicates PRESet ON.
MODE
Purpose:
CC, CR, CV or CP operating mode query command.
Command Syntax:
MODE?{NL}
Description:
MODE? returns the present operating mode status, ”0” indicates CC MODE, ”1” indicates CR
MODE, and “2” indicates CV MODE, “3” indicates CP MODE.