Sorensen SLD-Series DC Load

3-8

M540075-01 Rev C

3.1.2 Summary

Chart

The front panel controls are summarized in Figure 3-3.

CC: Constant Current
CR: Constant Resistance
CV: Constant Voltage

MODE

ON
OFF

PRESet

G/NG Volt Upper Limit
G/NG Volt Lower Limit
G/NG Current Upper Limit
G/NG Current Lower Limit

LIMIT

OFF
ON

Vsense

OFF
ON

G/NG

Load On Voltage
Load Off Voltage
iStart Current
iStop Current
iStep Current
Vthreshold

Normal
OCP

Test Mode

OFF
ON

Parallel Setup AIIb

CONFIG

DYNamic
STAtic

DYN/STA

ON
OFF

SHORT

ON
OFF

LOAD

High Level Current
Low Level Current
Time at High Level
Time at Low Level
Slew Rate

DYN STTING

SLD Dual DC Loads

Button Functions

Figure 3-3 Organization of Front Panel Controls of SLD Electronic Load

3.2

Operation – Setup and OCP Test

The SLD-series dual input DC loads can operate in five modes: Constant Current (CC),
Constant Voltage (CV), Constant Resistance (CR), Shorts, and Dynamic. Dynamic mode allows
controlled current transitions between two states in CC.

The following procedure is outlined in Figure 3-4. The number within parentheses following a
front panel key name, refers to the labels in Figure 3-1 and Figure 3-2

3.2.1

Overview of Front Panel Operation

Select channel A or B (2) to input settings. Next, set Limits (15) for the go/no go upper and
lower limits for voltage and current. Then configure (14) the remainder of the system
settings for V-sense control, load On voltage, load Off voltage, and OCP test setup.

Once you’ve set up the initial parameters, use the MODE key (3) to select CC, CR, or CV
mode and set up the applicable parameters for each. Select either Static mode (11) for any
of CC, CR, or CV, or Dynamic mode (11) for CC only. In Dynamic mode, use the Dynamic
Setting key (12) to set values for the required T

HIGH

, T

LOW

, and Slew parameters.

Finally, apply load to the input terminals with the Load key (10), which toggles the Load On
and Off.