Introduction

Sorensen SLD-Series DC Load

1-4

M540075-01 Rev C

CV Mode:

(Constant Voltage Mode)

At constant voltage mode; the SLD-series Electronic Load will attempt to sink enough
current until the load input voltage is equaled to the programmed value (see Figure 1-7).

I

V

LOAD
CURRENT

INPUT VOLTAGE

VOLTAGE
SETTING

Figure 1-7 Constant Voltage Mode

Dynamic Waveform Definition:

There are six parameters to generate dynamic waveform or pulse waveform: High Load
Level, Low Load Level, T

HIGH

, T

LOW

, and Rise and Fall Slew Rates. The SLD-series

Electronic Load will sink current from the power source proportional to the dynamic
waveform, the definition for which is shown in Figure 1-8. The period of dynamic
waveform is T-high + T-low.

Dynamic frequency = 1 /( T-high + T-low )
Duty cycle = T-high / ( T-high + T-low )


Figure 1-8 Dynamic Waveform