AMETEK SLD Series User Manual
Page 62
Remote Programming
Sorensen SLD-Series DC Load
5-8
M540075-01 Rev C
POWER Level
Purpose:
The load power setting in Constant Power mode.
Command Syntax:
CP:{HIGH|LOW}{SP}{NR2}{;|NL}
Description:
This command is used to set the load Power level of electronic load modules.
LOAD ON/OFF
Purpose:
Turn the Electronic load module input ON or OFF.
Command Syntax:
LOAD{SP}{0FF|ON}{NL}
Description:
This command sets the Electronic load to sink current from power source.
For Example :
LOAD OFF ; load are at input OFF condition.
LOAD ON ; Set load to LOAD ON status, this load is ready to sink current from the power source.
LOAD
ON VOLTAGE Setting
Purpose:
The Load ON voltage setting (Initial is 1.0V) of DC electronic load.
Command Syntax:
LDON{SP}{NR2}{;|NL}
Description:
The Load On voltage can be adjusted by the LDON command. The range is 0.1-25.0 V (Resol. =
0.1V). The load will start to sink current if power source output voltage is higher than Load On
voltage.
For Example:
LDON 2.5; Set the Load On voltage to 2.5V, The load will start to sink current when the power
source output voltage is higher than 2.5V.
LOAD OFF VOLTAGE Setting
Purpose
:
The Load OFF voltage setting (Initial is 0.5V) of DC electronic load.
Command Syntax:
LDOF{SP}{NR2}{;|NL}
Description:
The Load Off voltage can be adjusted by the LDOF command; the adjust range is 0.1-load on
voltage (Resol. = 0.1V). The load will stop to sink current if power source output voltage is lower
than Load Off voltage.
For Example :
LDOF 2.0 ; Set the Load Off voltage to 2.0V. The load will start to sink current when power source
output voltage is lower than 2.0V.