Applications

Sorensen SLD-Series DC Load

4-4

M540075-01 Rev C

4.3.2 Dynamic

Mode:

Built-in Pulse Generator

: (Fig 4-4)

Major application

:

Power supply load transient response testing, power recovery time testing, pulse load
simulation, power component testing

Description:
The maximum Rise/Fall current slew rate or minimum Rise/fall time is the time required for
the load input to change from 10% to 90% or from 90% to 10% of the programmed High to
Low load level.

• Rise slew rate = | Ilow - Ihigh | / Ta (A/us)
• Fall slew rate = (Ihigh - Ilow) / Tb (A/us)
• Rise time = Ta = | Ilow - Ihigh | / Rise slew rate
• Fall time = Tb = (Ihigh - Ilow) / Fall slew rate


Major application:

Simulate real load condition, Battery discharge testing

T

HIGH

T

LOW

T a

Tb

T

LOAD CURRENT

I

HIGH

I

LOW

Figure 4-4 Dynamic Load Current with Independent Programmed Rise/Fall Slew Rate