SYMPTOM CHECK ACTION

No output; all displays and
indicators are blank

Is the utility power present?

Connect power

Is the AC input power cord
defective?

Check continuity; replace
if necessary

Unit starts but there is no
output; OVP and FAULT
indicators are off

Is the AC input voltage within
the allowed range?

Connect to appropriate
voltage source

Output is present
momentarily but quickly
shuts off; OVP and FAULT
indicators are off

Does the voltage of the AC
source sag when load is
applied?

Ensure that the AC
source has low output
impedance and can
supply adequate current

No output; displays and
indicators are functional

Is the OVP indicator on?

See Section 3.5, OVP
Operation

Is the EXT–OFF indicator on? See Section 4.5,

EXTERNAL–OFF Control

Is the REM indicator on?

See Section 4.4, Remote
Programming

Is the FAULT indicator on?

Determine if an overtem-
perature condition exists;
allow unit to cool and
then toggle the OUTPUT
switch off and back on

If an overtemperature is
not present, an internal
fault exists

Output turns off as soon as
POWER switch is turned on;
OVP indicator turns on

Is remote sense selected,
but sense leads are not
connected?

See Section 2.10, Load
Voltage Sensing

Output voltage will not adjust Is the unit in current–mode?

See Section 3.4, Mode of
Operation Setup

Output current will not adjust Is the unit in voltage–mode?

See Section 3.4, Mode of
Operation Setup

Front panel controls are
nonfunctional

Is unit in remote control with
LOCAL–LOCKOUT enabled?

See Section 3.6.4,
LOCAL/REMOTE Switch

Front panel controls are
nonfunctional; voltage
display is off

Is slave mode selected?

See Section 2.12, Parallel
Operation

Large AC component
present in output

Is the power converter
unstable?

See Section 2.9.3, Noise
and Impedance Effects

Table 5–1. Troubleshooting Guide