4.4.6

Resistance Programming of OVP

Refer to Figure 4-8 for setting up resistance programming of OVP.

1.

Set Position-6 of the SETUP switch to OFF for 0–5 VDC programming range.

2.

Connect the external resistance, 0–5 k

, to the REMOTE ANALOG INTERFACE

connector, from Pin-17 to Pin-16.

3.

Connect a jumper from Pin-16 to Pin-3.

4.

Program the other parameters to the desired limit values: VOLTAGE
PROGRAMMING INPUT, Pin-9, and the CURRENT PROGRAMMING INPUT,
Pin-10, with respect to Pin-12. AUXILIARY 5VDC OUTPUT, Pin-15, could be
used for full–scale programming.

5.

Connect Pin-1, ANALOG–CONTROL, of the REMOTE ANALOG INTERFACE
connector to Pin-6.

Figure 4-8. Resistance Programming of OVP