7. Interrupting an evaporation process:
Press the left-hand function key "mode".
Select "Stop evaporation" and confirm.
Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:
At maximum speed: the stop valve closes the connection to the
vacuum pump, the aeration valve opens; the rotor speed is
maintained for 30 sec (aeration phase, to avoid delays in
boiling).
Rotor decelerates until it is at a standstill.
After the standstill of the rotor the lid can be opened in order to
remove the rotors.
Function and control elements:
1 Lid
2 Rotor chamber
3 Control panel
4 Mains power switch
5 Lid lock device
6 Rotor shaft
7 Name plate
Total view of the rotational vacuum concentrator
1 Electrical vacuum measuring
sensor connection
2 Option: Serial interface
3 Option: Serial interface
remote
4 Power supply connection of
vacuum pump
5 Power supply connection of
pressure control valve
6 Power supply connection of
the stop valve
7 Mains connection and mains
fuse protection
8 Equipotential bonding screw
9 Option: Valve block
10 Vacuum connection
Rear view of the rotational vacuum concentrator
10
1
2
3
4
5
6
7
1
2
3
4
5
6
7
8
9
10