8.3 Autoclaving
The service life of the accessories essentially depends on the
frequency of autoclaving and use.
Remove the existing O-rings so that the trapezoid notch can be
reached by the hot steam. This prevents the formation of humidity
in the notch..
Replace the accessories immediately when the parts show
changes in color or structure or in the occurrence of leaks etc.
Accessories
max. temp.
°C
min. time
min
max. time
min
Cycles
(max.)
Glass tubes
134-138
3
40
-
Polycarbonate tubes
115-118
30
40
20
Polypropylene tubes
115-118
30
40
30
Teflon tubes
134-138
3
5
100
Aluminium rotor
134-138
3
20
-
Aluminium rotor accessories
134-138
3
20
-
8.4 Service
In the event of service work that requires the removal of the panels,
there is a risk of electric shock or mechanical injury. Only qualified
specialist personnel is authorised to perform this service work.
The rotational vacuum concentrator is subject to high mechanical
stress. In order to be able to withstand this high level of stress, high-
quality components were used during the production of the rotational
vacuum concentrators. Nevertheless, wear cannot be excluded and it
may not be visible from the outside.
This is why we recommend having the rotational vacuum
concentrator checked by the manufacturer during an inspection once
per year.
Information and appointments:
in Germany:
use the service request form at
www.martinchrist.de
[Service Area]
or contact
Martin Christ Gefriertrocknungsanlagen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Tel. +49 (0) 55 22 / 50 07-8425
Fax +49 (0) 55 22 / 50 07-9425
E-mail: [email protected]
in other countries: