Martin Christ RVC 2-33 CDplus User Manual
Page 44

RVC 2-33 CD
&
Using the RVC
44
Version 01/2009, Rev. 1.21 of 25/11/2013 - sb
Translation of the original operating manual
Deleting a program
To delete a program:
Select "Delete program" from the program administration menu
(see fig. 6.18) and confirm.
Select a program from the program list and confirm. A message
will be displayed:
Fig. 6.26: Message for deleting a program
Confirm the message. The program is deleted.
Program memory information
Select "Program memory information" from the program
administration menu (see fig. 6.18) and confirm. Information
concerning free program locations and sections are displayed:
Fig. 6.27: Program memory information
6.3.5.3 Special Functions
Pressure increase test (only if a vacuum sensor is installed)
The pressure increase test can be used in order to determine whether
the evaporation in the rotor chamber is completed:
During the pressure increase test, the stop valve separates the rotor
chamber from the cooling trap or from the vacuum pump. The system
evaluates the increase in pressure in the chamber that is caused by
additional vapour being released.
To perform a pressure increase test, the limits for the increase in
pressure and the test time have to be defined.
To enter the limits for the increase in pressure and the test time:
Select "Special functions" from the main menu.
Open the menu "Pressure increase test" and select "Parameter
pressure increase test".