Martin Christ RVC 2-33 CDplus User Manual
Page 55
![background image](/manuals/739299/55/background.png)
RVC 2-33 CD
&
Maintenance and Service
Version 01/2009, Rev. 1.21 of 25/11/2013 - sb
55
Translation of the original operating manual
8.1.2 Rotor Chamber
If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in
order to prevent corrosion.
Cleaning of the rotor chamber:
1 Rotor chamber
2 Cap
3 Rotor bearing
Fig. 8.1: Rotor chamber with rotor shaft
Open the lid.
Disconnect the power cord from the wall outlet or instrument
receptacle before cleaning.
Hot surface!
Let the rotor chamber, the rotor shaft and the lid cool down to room
temperature!
Pull off the cap.
1 Allen screw
Fig. 8.2: Removing the allen screw
Remove the allen screw with the supplied allen wrench (size 5).
1
1
2
3