RVC 2-33 CD
&
Malfunctions and Error Correction
52
Version 01/2009, Rev. 1.21 of 25/11/2013 - sb
Translation of the original operating manual
7.1.4 Emergency Lid Release
In the event of a power failure, it is possible to manually open the lid
of the rotational vacuum concentrator.
Do not actuate the emergency lid release before the rotor has
stopped and the pressure has been relieved.
Do not use the emergency lid release to open the lid during
normal operation. Unlocked lid locks increase the risk of
accidents!
Switch off the mains power switch and disconnect the power cord
from the socket.
Check if the rotor is at a standstill and the rotor chamber is
completely aerated.
Pull the unit a bit over the edge of the table and jack it up (see
Fig. 7.1). Ensure that the centre of gravity of the unit is still on the
table. The unit must be secured by a second person.
Fig. 7.1: Position of the unit over the edge of the table
Unscrew the two black plastic screws (approx. 8 mm).
Remove the stoppers, e.g. with a screwdriver. The stoppers are
connected to a mechanical release string (see fig. 7.2).
1 Release string
2 Plastic screw
3 Stopper
Fig. 7.2: Release string with stopper and screw
2
3
1