3 evaporation summary – Martin Christ RVC 2-33 CDplus User Manual
Page 73

RVC 2-33 CD
&
Appendix
Version 01/2009, Rev. 1.21 of 25/11/2013 - sb
73
Translation of the original operating manual
11.3 Evaporation Summary
Vials
Number of
samples
Sample volume
(ml)
Temperature (°C)
Pump
Cooling trap
CT 02-50
Total time (min)
Water
100 ml test tube
12
50
30
MZ 2C
yes
1300
100 ml test tube
12
50
45
MZ 2C
yes
910
100 ml test tube
12
50
60
MZ 2C
yes
810
100 ml test tube
12
50
80
MZ 2C
yes
615
Acetonitrile C
2
H
3
N
100 ml test tube
12
50
30
MZ 2C
yes
230
100 ml test tube
12
50
45
MZ 2C
yes
160
100 ml test tube
12
50
60
MZ 2C
yes
160
100 ml test tube
36
1
30
MZ 2C
yes
90
1,4 Dioxane C
4
H
8
O
2
100 ml test tube
12
50
30
MZ 2C
yes
270
100 ml test tube
12
50
45
MZ 2C
yes
225
100 ml test tube
12
50
60
MZ 2C
yes
155
100 ml test tube
12
50
80
MZ 2C
yes
110
Toluol C
6
H
5
CH
3
100 ml test tube
12
50
30
MZ 2C
yes
165
100 ml test tube
12
50
45
MZ 2C
yes
150
100 ml test tube
12
50
60
MZ 2C
yes
110
100 ml test tube
12
50
80
MZ 2C
yes
95
Methanol CH
3
OH
100 ml test tube
12
50
30
MZ 2C
yes
285
100 ml test tube
12
50
45
MZ 2C
yes
210
100 ml test tube
12
50
60
MZ 2C
yes
195
100 ml test tube
12
50
80
MZ 2C
yes
180
Ethanol C
2
H
6
O
100 ml test tube
12
50
30
MZ 2C
yes
235
100 ml test tube
12
50
45
MZ 2C
yes
197
100 ml test tube
12
50
60
MZ 2C
yes
180
100 ml test tube
12
50
80
MZ 2C
yes
145