In this example, the instrument is set up to log samples only if the temperature of the sample is
above 18
°
C.
To alter the switch so that data is logged when the specific conductance is greater than 100 uS/cm
and temperature is greater than 18
°
C, first press 5 for SpCond mS/cm.
The prompt 0-lower or 1-higher than (1) will appear on the screen under the menu.
First type 1 (higher than) and press Enter.
The screen prompt will ask for the condition: Threshold (0.000):
Enter your new condition in uS/cm (100) and press Enter.
The screen will now display the altered menu which shows the new condition and data will be
logged only when the temperature is greater than 18
°
C AND the specific conductance is greater
than 100 uS/cm.
========================================
Set-Condition
----------------------------------------
1-(*) Temp C
F-( ) DO %
2-( ) Temp F
G-( ) DO mg/L
3-( ) Temp K
H-( ) Depth ft
4-( ) SpCond mS/cm
I-( ) Depth m
5-( ) SpCond uS/cm
J-( ) pH
6-( ) Cond mS/cm
K-( ) pH mV
7-( ) Cond uS/cm
L-( ) ORP mV
8-( ) Resist MOhm*cm
M-( ) NH4+ mg-N/L
9-( ) Resist KOhm*cm
N-( ) NH3 mg-N/L
A-( ) Resist Ohm*cm
O-( ) NO3- mg-N/L
B-( ) TDS g/L
P-( ) Turb NTU
C-( ) TDS kg/L
Q-( ) TDG mm-Hg
D-( ) TDS mg/L
R-( ) Batt V
E-( ) Sal ppt
Condition: Temp C > 18.000
Select option (0 for previous menu):