WARNING: Low RTC battery
Lithium battery that powers the real time clock is low.
Replace lithium battery.
ERROR(000): Undetermined
Internal error. The system is unable to determine the cause.
Repeat action. If necessary reset by removing the battery lid
and repeat action.
ERROR(001): Illegal input conversion
Internal error. The system encountered an unknown
character in the last I/O conversion. Repeat action.
ERROR(002): Conversion overflow
Input number exceeds system limit.
ERROR(003): Input error
User input error.
ERROR(004): Input out of range
Input number exceeds allowable range.
ERROR(005): Illegal hex nibble
System encountered bad data during system software update.
ERROR(006): Illegal address
System encountered bad address during system software
update.
ERROR(007): Illegal record
System encountered bad record during system software
update.
ERROR(008): Illegal page number
System encountered bad code page during system software
update.
ERROR(009): Mismatch hex checksum
System has detected bad checksum during system software
update.
ERROR(010): Hex load abort
System has detected abort sequence during system software
update.
ERROR(011): SFLS erase suspend
Internal error. System software erasure was suspended.
ERROR(012): SFLS erase fail
Fatal hardware failure. Unable to remove old code during
system software update. Battery is low or program failed to
save system setup information.
ERROR(013): SFLS program fail
Fatal hardware failure. Failed to update system software or
failed to save system setup.
ERROR(014): SFLS VPP low
Fatal hardware failure. Unable to raise programming voltage
or low battery voltage.
ERROR(015): SFLS command sequence
Internal error. Bad programming sequence during system
software update.
ERROR(016): Write to occupied sflscell
Internal error. Hardware failure during system software
update.
ERROR(017): SC timeout
Internal error.
ERROR(018): SQ empty
Internal error.
ERROR(019): SQ full
Internal error.
ERROR(020): Rx overrun
Internal error.
ERROR(021): Rx framming
Internal error.
ERROR(022): Rx parity
Internal error.
WARNING: High DO charge
DO charge current exceeds limit (25 to 75 during calibration.
Probe maintenance is required.
WARNING: Out of range
Calibration procedures determine that actual solution and
user inputted number cannot be resolved.
ERROR: Sensor disabled
System is not configured with appropriate sensor for
calibration.