YSI 6000UPG User Manual
Page 131

9-3
Master: aM!
Initiate a measurement.
6000
UPG
: atttn[CR][LF]
ttt - Maximum time in seconds the 6000
UPG
will take to complete the measurement.
n - Number of data that will be available when the measurement is completed. This number is the
same as the number of output parameters set in the Report-setup menus (see Section 4.6). Date and
time are also counted. If the number of parameter set is greater than 9, it is truncated to 9.
After finishing the measurement, the 6000
UPG
will send a service request "a[CR][LF]" to the bus
master. The bus master can then retrieve the measurement result by "D0" to "D9" commands (see
below). If the 6000
UPG
does not send a service request within the specified maximum time, the
measurement was canceled. The bus master can then send a "V" command (see below) to obtain
the specific error code or restart with another "M" command.
Master:
aV!
Initiate a verifying sequence.
6000
UPG
: atttn[CR][LF]
ttt - The time the 6000
UPG
will take to execute the verifying sequence. This is always 004 seconds.
n - Number of data from the verifying sequence. This is always 5. The 5 data that will be available
are date, time, battery voltage, free flash disk space in KB, and the last system error code (in this
order).
After finishing the verifying sequence, the 6000
UPG
will send a service request "a[CR][LF]" to the
bus master. The bus master can then retrieve all verifying data with a single "D0" command (see
below).
Master: aI!
Send identification.
6000
UPG
: allccccccccmmmmmmvvvxxx...xxx[CR][LF]
l - 2 character SDI-12 level number. For version 1.00 of the EMS6000 software, this field is "10".
c - 8 character manufacturer identification. This field always contains "YSIIWQSG" (YSI Inc.,
Water Quality Systems Group).