Fad-30en, With ethernet interface, Front view – Flintec FAD-30 User Manual
Page 40: With, Thernet, Nterface
FAD-30 Technical Manual, Rev. 1.06 December 2012
Page 40 of 48
9. FAD-30EN
–
WITH
E
THERNET
I
NTERFACE
FAD-30EN instruments are state-of-the-art strain gauge load cell signal digitizers with Ethernet interface. These
instruments are used for any type of process weighing and force measurement including tank and silo
weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc.
9.1. Front View
There are 3 status LEDs on the front panel which indicate the operational instrument status (Refer to table 9.1).
The setup switch on front panel of the instrument is used for diagnostics (Refer to chapter 12)
When the error LED is ON, the other two LED indicate the error type (Refer to chapter 11 for details).
Load cell connection, Ethernet, power supply and serial interface terminals are located at the front of the DIN
rail mount instrument (See figure 9.1).
Status LEDs
Ethernet Terminal
Load Cell Connector
Setup Switch
Power Supply Terminal
RS-232C Terminal
Figure 9.1
– Front view
The meanings of these LEDs in operation are described below.
LED
Operational Mode
Symbol
Name
Weight / Force
Count
Stable
Stable
Unstable
Off for 0.3 seconds in 2 seconds
period (No stable indication)
Centre
of Zero
in the centre of zero range
( -0.25 e < w < 0.25 e )
Out of centre of zero range
Always off
( No centre of zero indication )
Err
Error
( * )
ADC conversion error
Digital processing error
No error
ADC conversion error
Digital processing error
No error
Off
On
Flashing Off for 0.3 seconds
( * ) : Refer to the error table in chapter 11
Table 9.1
– Status LEDs