FAD-30 Technical Manual, Rev. 1.06 December 2012

Page 35 of 48

ST Status LED

State

Indication

Comment

Off

No power or not initialized

No power or profibus module is in initializing

Green

Initialized

Flashing Green

Initialized, diagnostic event(s) present

Diagnostic is active

Red

Exception error

There is an exception error

OP Operation Mode LED

State

Indication

Comment

Off

Not on-line / No power

Check power and cable

Green

On-line, data exchange

-

Flashing Green

On-line, clear

-

Flashing Red (2x)

PROFIBUS configuration error

Check GSD file configuration.

7.2. Electrical Connections

Profibus and RS-232C connections are shown in figure 7.2.

PROFIBUS Connector (DB9F)

Pin

Signal

Description

1

-

-

2

-

-

3

B Line

Positive RxD / TxD, RS-485 level

4

RTS

Request to send

5

GND Bus

Ground (isolated)

6

+5V Bus output

+5V termination power (isolated)

7

-

-

8

A Line

Negative RxD / TxD, RS-485 level

9

-

-

Housing

Cable shield

Ground

Figure 7.2

– FAD-30PB interface connections

RS-232C Serial Interface

Use

Setup via xFace

Baud rate

9600 bps

Length and parity

8 bit no parity

Start / Stop bits

1 start bit and 1stop bit

Profibus-DP Interface

Use

Interfacing with PC or PLC

Data format

Profibus

Baud rate

Automatically detected and supported baud rates are 9.6 kbps, 19.2 kbps, 45.45
kbps, 93.75 kbps, 187.5 kbps, 500 kbps, 1.5Mbps, 3 Mbps, 6 Mbps and 12
Mbps. No „baud rate‟ instance exists.

Warning: Connect the shield to the reference ground or shield pin of the power connector.
Warning: Disconnect xFace PC software for Profibus interfacing.