beautypg.com

A/d converter setup and calibration, Scale parameters, Figure 5.3 – fad-30 setup parameters – Flintec FAD-30 User Manual

Page 13: Increased external resolution, Digital filter

background image

FAD-30 Technical Manual, Rev. 1.06 December 2012

Page 13 of 48

5.3. A/D Converter Setup and Calibration

WARNING: Read this manual carefully before you setup and calibrate the instrument. This will increase the
performance of your weighing system by applying proper setup and calibration.

Visual Weight Display: Displays the weight, count or force value

Toolbox: Contains shortcuts of some special commands

A/D Converter Parameters:

This block allows the user to setup the A/D Converter‟s parameter

related to the operation mode. Refer to chapter 5.3.1

RS-232C: RS-232C serial port communication settings (not for FAD-30 and FAD-30MB)

RS-485: RS-485 port communication settings. (FAD-30 and FAD-30MB only)

Description: This block provides some clear-text explanations

Read from A/D Converter: Click this button to read the parameter settings from the instrument
Write to A/D Converter: Click this button to save the parameter settings to the instrument
Default: Click this button to load the factory default settings
Clear: Empties the parameter settings

Figure 5.3

– FAD-30 Setup Parameters

For entering the parameter values enter the setup tab, adjust the parameter settings and then write this data to
the A/D Converter. After changing parameter settings the instrument always requires a re-calibration.

5.3.1.

Scale Parameters

In the setup menu the scale parameters of FAD-30 can be viewed, changed or saved to the instrument. These
parameter settings are:

Increased External Resolution

For service purposes this parameter enables the 10 times higher resolution of the weight value than the defined
interval in weighing and force mode. Increased resolution must be disabled for use in normal operation (Not
available in Count Mode). Default setting: „Disable‟.

Digital Filter

Environmental noise like bounding forces, air flow, vibration, motor control instruments etc. may disturb the load
cell signal. The selection of proper filter settings determines how quickly the system will react to the load cell
signal.**Load cell signal digital filtering is done according to this parameter settings. The settings can be
changed from 0 (fastest settling) to 9 (slowest set

tling). Default setting: „7‟.

Setting

0

1

2

3

4

5

6

7

8

9

Values/s

1600

800

400

200

100

100

100

50

50

50

Settling time [ms]

80

140

250

180

300

420

380

620

720

1800