FAD-30 Technical Manual, Rev. 1.06 December 2012

Page 37 of 48

8. FAD-30PN

WITH

P

ROFI

N

ET

I

NTERFACE

FAD-30PN instruments are state-of-the-art strain gauge load cell signal digitizers with ProfiNet interface. These
instruments are used for any type of process weighing and force measurement including tank and silo
weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc.
The EDS file is available on a CD which is supplied together with the instrument.

8.1. Front View

There are 6 status LEDs on the front panel which indicate the operational instrument status (Refer to table 8.1).
and the ProfiNet status. The setup switch on front panel of the instrument is used for ProfiNet addressing
without PC (Refer to chapter 5.6) and for diagnostics (Refer to chapter 12).
When the error LED is ON, the other two LED indicate the error type (Refer to chapter 11 for details).
Load cell connection, ProfiNet connection, power supply and serial interface terminals are located at the front of
the DIN rail mount instrument (See figure 8.1).

Status LEDs

ProfiNet Connector

Load Cell Connector

Setup Switch

Power Supply Terminal
RS-232C Terminal

Figure 8.1

– Front view

The meanings of these LEDs in operation are described below.

LED

Operational Mode

Symbol

Name

Weight / Force

Count

Stable

Stable
Unstable

Off for 0.3 seconds in 2 seconds

period (No stable indication)

Centre

of Zero

in the centre of zero range

( -0.25 e < w < 0.25 e )

Out of centre of zero range

Always off

( No centre of zero indication )

Err

Error

( * )

ADC conversion error

Digital processing error

No error

ADC conversion error

Digital processing error

No error

Off

On

Flashing Off for 0.3 seconds


( * ) : Refer to the error table in chapter 11.

Table 8.1

– Status LEDs