Flintec FAD-30 User Manual
Page 23
FAD-30 Technical Manual, Rev. 1.06 December 2012
Page 23 of 48
5.10. Modbus Data Structure (for FAD-30MB + FAD-30EN only)
Address R/W
Word Command
Definition
40001
R
2
Weight / Force / Count Data
40003
R
1
Status
D0
0
– System Ready
1
– System Busy
D1
0
– Error
1
– Data ok
D2
0
– Weight Stable
1
– Weight unstable
D3
0
– Gross Mode
1
– Net mode
D4
Not Used
D5
0
– Weight / Force
1
– Count Mode
D6-D11
Not Used
D12
0
– Out of zero range 1 – Weight is in zero range
D13
D14
D15
Error Code
0
No Errors
1
ADC out of range
2
ADC over range
3
ADC under range
4
System error
5
In setup mode
6
Low/High voltage det.
40004
R
2
Tare weight
40006
R
2
Gross weight
40008
R
1
Status
Motion, Net mode, Data ok, (image of register 40003)
40009
R/W
1
Control
0
None
1
Zero
2
Tare
3
Clear
40010
R/W
1
Calibration
0
None
188
Adjust Zero Calibration
220
Adjust Span Calibration (First load calibration weight
on scale and load 40011 with span test weight value)
40011
R/W
2
Span Calibration Value
40013
R
1
Calibration
Status
D0 .. D7
Calibration
Process
Status
1
Ready for calibration
3
Zero calibration in process
4
Span calibration in process
9
Error (Refer to D8 ... D15 )
D8 .. D15
Calibration
Errors
1
Calibration Timeout
- Restart calibration
2
ADC Error
- Re-energize the instrument
- If seen again, change the board.
3
Instrument cannot be calibrated
- Check load cell cable
- Re-energize the instrument
34
Instrument cannot be calibrated
- Load cell signal is very low or too high
35
Calibration Error
- Calibration test weight is not enough
- Increase calibration weight value (40011)
- Check load cell connections
37
Scale unstable
- Wait until scale become stable
- Check grounding wiring
40014
R/W
1
Operation Mode
Selector
0
Count Mode Unipolar
1
Count Mode Bipolar
2
Force Mode Unipolar
3
Force Mode Bipolar
4
Weight Mode Unipolar