FAD-30 Technical Manual, Rev. 1.06 December 2012

Page 23 of 48

5.10. Modbus Data Structure (for FAD-30MB + FAD-30EN only)


Address R/W

Word Command

Definition

40001

R

2

Weight / Force / Count Data

40003

R

1

Status

D0

0

– System Ready

1

– System Busy

D1

0

– Error

1

– Data ok

D2

0

– Weight Stable

1

– Weight unstable

D3

0

– Gross Mode

1

– Net mode

D4

Not Used

D5

0

– Weight / Force

1

– Count Mode

D6-D11

Not Used

D12

0

– Out of zero range 1 – Weight is in zero range

D13
D14
D15

Error Code

0

No Errors

1

ADC out of range

2

ADC over range

3

ADC under range

4

System error

5

In setup mode

6

Low/High voltage det.

40004

R

2

Tare weight

40006

R

2

Gross weight

40008

R

1

Status

Motion, Net mode, Data ok, (image of register 40003)

40009

R/W

1

Control

0

None

1

Zero

2

Tare

3

Clear

40010

R/W

1

Calibration

0

None

188

Adjust Zero Calibration

220

Adjust Span Calibration (First load calibration weight
on scale and load 40011 with span test weight value)

40011

R/W

2

Span Calibration Value

40013

R

1

Calibration
Status

D0 .. D7
Calibration
Process
Status

1

Ready for calibration

3

Zero calibration in process

4

Span calibration in process

9

Error (Refer to D8 ... D15 )

D8 .. D15
Calibration
Errors

1

Calibration Timeout
- Restart calibration

2

ADC Error
- Re-energize the instrument
- If seen again, change the board.

3

Instrument cannot be calibrated
- Check load cell cable
- Re-energize the instrument

34

Instrument cannot be calibrated
- Load cell signal is very low or too high

35

Calibration Error
- Calibration test weight is not enough
- Increase calibration weight value (40011)
- Check load cell connections

37

Scale unstable
- Wait until scale become stable
- Check grounding wiring

40014

R/W

1

Operation Mode
Selector

0

Count Mode Unipolar

1

Count Mode Bipolar

2

Force Mode Unipolar

3

Force Mode Bipolar

4

Weight Mode Unipolar