Calibration status – Flintec FAD-30 User Manual
Page 29
![background image](/manuals/676053/29/background.png)
FAD-30 Technical Manual, Rev. 1.06 December 2012
Page 29 of 48
Calibration Status
Bit no.
T_DW2 descriptions when read command is ‘Calibration Status’.
Refer to PLC Output to FAD-30CO Input R_DW3
D31…D16
Not in use
D15…D8
Calibration
error
0000 0001
Calibration timeout: - Restart calibration
0000 0010
ADC error: - Re-energize the instrument
0000 0011
Instrument cannot be calibrated
- Check load cell cable
- Re-energize the instrument
0010 0010
Instrument cannot be calibrated:
- Load cell signal is very low or too high
0010 0011
Calibration Error
- Calibration test weight is too small; - Increase calibration
weight value (Write test weight value from PLC Output to FAD-
30CO Input R_DW2 then restart the calibration)
- Check load cell connections
0010 0101
Scale unstable
- Wait until scale becomes stable
- Check ground wiring
D7…D0
Calibration
process
status
0000 0001
System ready for calibration
0000 0011
Zero calibration in process
0000 0100
Span calibration in process
0000 1001
Error (Calibration error)