Calibration status – Flintec FAD-30 User Manual

Page 29

background image

FAD-30 Technical Manual, Rev. 1.06 December 2012

Page 29 of 48

Calibration Status

Bit no.

T_DW2 descriptions when read command is ‘Calibration Status’.
Refer to PLC Output to FAD-30CO Input R_DW3

D31…D16

Not in use

D15…D8

Calibration
error

0000 0001

Calibration timeout: - Restart calibration

0000 0010

ADC error: - Re-energize the instrument

0000 0011

Instrument cannot be calibrated
- Check load cell cable
- Re-energize the instrument

0010 0010

Instrument cannot be calibrated:
- Load cell signal is very low or too high

0010 0011

Calibration Error
- Calibration test weight is too small; - Increase calibration
weight value (Write test weight value from PLC Output to FAD-
30CO Input R_DW2 then restart the calibration)
- Check load cell connections

0010 0101

Scale unstable
- Wait until scale becomes stable
- Check ground wiring

D7…D0

Calibration
process
status

0000 0001

System ready for calibration

0000 0011

Zero calibration in process

0000 0100

Span calibration in process

0000 1001

Error (Calibration error)